Tel:+86-535-2169175
Mobile:+86-535-2169072
E-mail:sales@brightstonemems.cn
Add.:No. 46, Taipei North Road, Development Zone, Yantai City, Shandong Province, China
Applications: Ventilator, oxygen chamber, incubator, oxygen collector, inert gas treatment, fermentation equipment, food detection, monitoring outdated growth, wine cellar gas storage, refinery diving.
Features
Zirconia sensitive test unit
Long service life, non-consumable testing technology
No need for reference gases
A varity of output siginal types are optional
High measurment accuracy
Internal integrated heating
Immediate calibration in the air
No need for stable external ambient temperature
Main Applications
Ventilator, oxygen chamber, incubator, oxygen collector, inert gas treatment, fermentation equipment, food detection, monitoring outdated growth, wine cellar gas storage, refinery diving.
Specifications
Heater voltage | Standard sensor | 4VDC±0.1VDC |
fast response sensor | 4.35VDC±0.1VDC | |
test wait mode | 2VDC±0.1VDC | |
@700℃ Oxygen pump impedance | <6kΩ | |
Operating temperature | -100℃~250℃ | |
Oxygen pressure monitoring range | 2mBar~3Bar | |
Test accuracy (oxygen partial pressure) | 5mBar | |
Response time (T90) | Standard sensor | <15s |
Fast response sensor | <4s | |
Start-up time | Cold start time | 60s |
Wait mode start | 20s | |
Test stability time | 180s | |
Weight | Standard sensor | 7g |
Fast response sensor | 9g | |
Detection velocity range | 1~10m/s |
- Prev:no